Remote Control
R&S
®
ESR
836User Manual 1175.7068.02 ─ 12
Example:
SWE:TYPE FFT
Selects FFT sweeps.
Manual operation: See "Sweep" on page 359
See "FFT" on page 359
See "Auto" on page 359
[SENSe:]SWEep:TYPE:USED?
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type> SWE | FFT
SWE
Normal sweep
FFT
FFT mode
*RST: SWE
Example:
SWE:TYPE:USED?
Usage: Query only
11.5.2.5 Trigger Configuration
The following commands configure triggered or gated measurements.
Triggering Measurements
TRIGger<n>[:SEQuence]:HOLDoff[:TIME]........................................................................836
TRIGger<n>[:SEQuence]:IFPower:HOLDoff..................................................................... 837
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis................................................................. 837
TRIGger<n>[:SEQuence]:LEVel[:EXTernal].......................................................................837
TRIGger<n>[:SEQuence]:LEVel:IFPower..........................................................................838
TRIGger<n>[:SEQuence]:LEVel:RFPower........................................................................ 838
TRIGger<n>[:SEQuence]:LEVel:VIDeo.............................................................................838
TRIGger<n>[:SEQuence]:SLOPe.....................................................................................839
TRIGger<n>[:SEQuence]:SOURce.................................................................................. 839
TRIGger<n>[:SEQuence]:TIME:RINTerval........................................................................ 840
TRIGger<n>[:SEQuence]:HOLDoff[:TIME] <Delay>
This command defines the length of the trigger delay.
A negative delay time (pretrigger) can be set in zero span only.
Suffix:
<n>
.
irrelevant