791User Manual 1175.7068.02 ─ 12
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
Manual operation: See "TOI" on page 224
See "TOI" on page 335
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
This command queries the results for the third order intercept point measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Return values:
<TOI> Third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES?
Outputs the measured value.
Usage: Query only
Manual operation: See "TOI" on page 224
See "TOI" on page 335
11.5.1.11 Measuring Statistics
The following commands configure and perform statistics measurements.
Remote Commands in Spectrum Analyzer Mode

Table of Contents