Remote Control
R&S
®
ESR
749User Manual 1175.7068.02 ─ 12
●
[SENSe:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel<channel>]
on page 731
●
[SENSe:]POWer:ACHannel:PRESet on page 725
●
[SENSe:]POWer:ACHannel:PRESet:RLEVel on page 726
●
TRACe<n>:DATA on page 726
[SENSe:]POWer:BANDwidth|BWIDth...............................................................................749
[SENSe:]POWer:BANDwidth|BWIDth <Percentage>
This command defines the percentage of the power with respect to the total power.
This value is the basis for the occupied bandwidth measurement (see [SENSe:
]POWer:ACHannel:PRESet on page 725).
Parameters:
<Percentage> 10 to 99.9PCT
*RST: 99PCT
Example:
POW:BWID 95PCT
Manual operation: See "% Power Bandwidth (span > 0)" on page 255
11.5.1.4 Measuring Spurious Emissions
The following commands configure and perform spurious emission measurements.
● Controlling the Measurement................................................................................749
● Managing a Sweep List.........................................................................................750
● Configuring List Evaluation................................................................................... 758
● Retrieving Results.................................................................................................759
Controlling the Measurement
Note that the spurious measurement must be initialized before you can start configur-
ing the sweep list or list evaluation.
Commands useful to retrieve results described elsewhere:
●
[SENSe:]SWEep:MODE on page 761
INITiate<n>:SPURious....................................................................................................749
MMEMory:STORe<n>:SPURious.................................................................................... 750
INITiate<n>:SPURious
This command initiates a Spurious Emissions measurement.
Suffix:
<n>
.
irrelevant
Example:
INIT:SPUR
Starts a Spurious Emissions measurement.
Usage: Event
Remote Commands in Spectrum Analyzer Mode