Spectrum Measurements
R&S
®
ESR
374User Manual 1175.7068.02 ─ 12
This softkey is available in single sweep mode.
Remote command:
CALCulate<n>:SGRam:FRAMe:COUNt on page 857
Spectrogram Clear
For spectrogram measurements only.
Resets the Spectrogram result display and clears its history buffer.
Remote command:
CALCulate<n>:SGRam:CLEar[:IMMediate] on page 649
6.2.7.2 Specifying the Sweep Settings
1. Press the Sweep Count softkey and enter the sweep count.
2. Press the Sweeptime Manual or Sweeptime Auto softkey to set the sweep time.
3. Press the Sweep Type softkey to select the sweep type.
4. Press the Sweep Points softkey and enter the number of sweep points.
5. Press the Continuous Sweep or Single Sweep softkey to select the sweep mode.
6. Press the Continue Single Sweep softkey to repeat the single sweep.
6.2.8 Triggering the Sweep – TRIG Key
The [TRIG] key is used to select trigger mode, trigger threshold, trigger delay, trigger
polarity and for gated sweep the gate configuration.
To open the Trigger menu
●
Press the [TRIG] key.
The "Trigger" menu is displayed.
Menu and softkey description
●
Chapter 6.2.8.1, "Softkeys of the Trigger Menu", on page 374
Tasks
●
Chapter 6.2.8.2, "Specifying the Trigger Settings", on page 381
●
Chapter 6.2.8.3, "Using Gated Sweep Operation", on page 381
6.2.8.1 Softkeys of the Trigger Menu
The following table shows all softkeys available in the "Trigger" menu. It is possible that
your instrument configuration does not provide all softkeys. If a softkey is only available
Configuration