Remote Control
R&S
®
ESR
673User Manual 1175.7068.02 ─ 12
11.4.6 Test Automation
General Scan Configuration..................................................................................673
Scan Table............................................................................................................ 674
Peak Search..........................................................................................................678
Peak Lists..............................................................................................................681
Trace Configuration...............................................................................................682
LISN Settings........................................................................................................ 684
11.4.6.1 General Scan Configuration
[SENSe:]SWEep:COUNt.................................................................................................673
[SENSe:]SWEep:COUNt:CURRent..................................................................................673
TRACe<n>:FEED:CONTrol<t>........................................................................................ 674
TRACe<n>:POINts.........................................................................................................674
[SENSe:]SWEep:COUNt <SweepCount>
This command defines the number of scans performed during a single sweep.
Parameters:
<SweepCount> Range: 0 to 32767
*RST: 0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single-sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation: See "Scan Count" on page 189
[SENSe:]SWEep:COUNt:CURRent
This command queries the number of scans that have started in single scan mode with
more than one scan (SENSe:SWEep:COUNt > 1).
Example:
SWE:COUNt 64
sets scan count to 64
INIT:CONT OFF
switches to single scan mode
INIT
starts a scan (without waiting for the scan end!)
SWE:COUN:CURR?
queries the number of started scans
Manual operation: See "Scan Count" on page 189
Remote Commands in Receiver Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals