Remote Control
R&S
®
ESR
697User Manual 1175.7068.02 ─ 12
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Outputs the measured value of marker 2.
In I/Q Analyzer mode, for "Real/Imag (I/Q)", for example:
1.852719887E-011,0
Manual operation: See "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 201
11.4.7.3
Using Delta Markers
CALCulate<n>:DELTamarker<m>:AOFF...........................................................................697
CALCulate<n>:DELTamarker<m>:LINK............................................................................697
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT........................................................... 698
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT...........................................................698
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK].........................................................699
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt.......................................................... 699
CALCulate<n>:DELTamarker<m>:MINimum:LEFT............................................................ 699
CALCulate<n>:DELTamarker<m>:MINimum:NEXT............................................................700
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK].......................................................... 700
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt........................................................... 700
CALCulate<n>:DELTamarker<m>:MODE......................................................................... 701
CALCulate<n>:DELTamarker<m>[:STATe]........................................................................ 701
CALCulate<n>:DELTamarker<m>:TRACe.........................................................................701
CALCulate<n>:DELTamarker<m>:X................................................................................. 702
CALCulate<n>:DELTamarker<m>:X:RELative?................................................................. 702
CALCulate<n>:DELTamarker<m>:Y................................................................................. 703
CALCulate<n>:DELTamarker<m>:AOFF
This command turns all active delta markers off.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT:AOFF
Switches off all delta markers.
Manual operation: See "Marker Configuration" on page 303
CALCulate<n>:DELTamarker<m>:LINK <State>
This command links delta marker 1 to marker 1.
If you change the horizontal position of the marker, so does the delta marker.
Remote Commands in Receiver Mode