Remote Control
R&S
®
ESR
861User Manual 1175.7068.02 ─ 12
CALCulate<n>:TLINe<Line> <Time>
This command defines the position of a time line.
Suffix:
<n>
.
irrelevant
<Line> 1 | 2
Selects the time line.
Parameters:
<Time> Range: 0 to 30000000000
*RST: (STATe to OFF)
Default unit: s
Example:
CALC:TLIN 10ms
Manual operation: See "Time Line 1 / Time Line 2" on page 213
CALCulate<n>:TLINe<Line>:STATe
<State>
This command turns a time line on and off.
Suffix:
<n>
.
irrelevant
<Line> 1 | 2
Selects the time line.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:TLIN2:STAT ON
Manual operation: See "Time Line 1 / Time Line 2" on page 213
Using Limit Lines
For commands that control limit lines see Chapter 11.4.8.3, "Using Limit Lines",
on page 705.
11.5.3.4 Markers
● Using Markers.......................................................................................................861
● Using Delta Markers..............................................................................................866
● Controlling Fixed Reference Markers....................................................................867
● Using Marker Functions........................................................................................ 869
Using Markers
Commands useful to control markers described elsewhere:
●
CALCulate<n>:MARKer<m>:AOFF on page 691
●
CALCulate<n>:MARKer<m>:FUNCtion:CENTer on page 688
Remote Commands in Spectrum Analyzer Mode