HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:99
NOTE!
The blood pump is preset to
stop when the priming
detector detects blood in the
venous blood line and the
blood path in the flow
diagram lights up.
4.2.13 Start the treatment
Before you begin
CAUTION!
Check that the blood path of the flow diagram lights up. If it does not light up, check
that the venous blood line is correctly placed in the priming detector. If the blood
path does not light up, the machine considers the patient not connected and as a
consequence the UF system is not controlling the patient ultrafiltration. If the air
detector has not been activated during priming it will be automatically activated
when the priming volume is achieved.
CAUTION!
All parameters and alarm limits must be suitable for the patientās needs and
tolerance.
CAUTION!
Additional measures to supervise the patient weight loss is recommended when
treating low weight patients or when performing long treatments. For Ultrafiltration
details, refer to Technical data and specifications.
Procedure
1) Check that the blood path is lit and that
treatment time is counting down. If this is
not the case, check that the venous blood
line is correctly placed in the priming
detector.