B:164 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
595 Total set UF volume
595 Total set UF volumeī¼
To obtain total UF volume start isolated UF.ī¼
Appears:
When there are 2 minutes left of a diffusion phase.
Machine actions:
None.
Operator action:
The dialysis machine calculates not to reach the intended
UF volume within the remaining treatment time. Do one of
the following things:
ā Start Isolated UF.
ā Adjust Isolated UF time.
ā Adjust Treatment time.
Mute time:
Permanent.