B:172 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
603 UF volume is set to 0.0 L
603 UF volume is set to 0.0 L
To accept press Confirm.
Appears:
When the priming detector detects blood and the set UF
volume is zero.
Machine actions:
Alarm sound is delayed for 2 minutes.
Operator action:
Press Confirm to confirm no ultrafiltration during treatment.
Mute time:
2 minutes.