B:138 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
564 Incorrect acidic concentrate
564 Incorrect acidic concentrate
Check set value and connected concentrate.
Appears:
Before treatment when the selected concentrate does not
match the concentrate connected to the machine.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
Check what concentrate is set and what is connected.
Make sure they are the same. See Section 4.2.3 ā€œSet up
the dialysis machineā€ on pageī™›A:77 in Operators
handbook.
Mute time:
2 minutes.