B:36 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
119 Low systolic blood pressure
119 Low systolic blood pressure
To clear the alarm press Confirm.
Appears:
When the patient's systolic blood pressure reaches the set
low alarm limit.
Machine actions:
None.
Operator action:
1. Check the blood pressure history:
- Press the BPM button.
- Select the History tab.
- Press History.
- Observe the measurement results from the treatment.
2. Do standard patient checks and take measures for
patients with circulatory disturbances.
3. If desired, measure the blood pressure manually, see
Section 8.3 ā€œDirect blood pressure measuringā€ on
pageī™›A:130 in Operators handbook.
4. If possible, adjust the alarm limits to the patient’s status:
- Press the BPM button.
- Select the Alarm tab.
- Press Set limits.
- Press Systolic Low and adjust the alarm limit.
5. Press OK.