HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions B:85
510 Blood pump overloaded
510 Blood pump overloadedī¼
Check blood pump for obstruction. To continue
press Confirm.ī¼
Appears:
When the blood pump is overloaded and can hardly run in
functional check.
Machine actions:
The blood pump stops.
Operator action:
1. Check that the blood pump segment is correctly placed,
see Section 4.2.4 āAttach the arterial blood lineā on
pageīA:79 in Operators handbook.
2. Press Confirm to continue.
If the attention appears again, contact an authorised
technician.
Mute time:
Permanent.