AK 98 Dialysis Machine - Attentions B:169
600 Profiling is not restarted
600 Profiling is not restarted
Check UF rate settings and press Confirm.
Appears:
When treatment is resumed after the dialyzer and the blood
lines have been changed during treatment.
Machine actions:
None.
Operator action:
1. Press the Fluid button.
2. Select the UF rate tab.
3. Press Profiling.
4. Check that the start and stop values for UF rate are the
correct ones and then press Confirm. If they are not
correct, deactivate profiling.
Change the values and reactivate profiling. See Section
7 ā€œProfilingā€ on pageī™›A:117 in Operators handbook.
Mute time:
-