B:46 AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
203 Heparin pump is overloaded
203 Heparin pump is overloaded
Check heparin line for obstruction.
Appears:
When the pressure in the syringe is too high.
Machine actions:
The heparin pump stops.
Operator action:
1. Check that the heparin syringe is installed.
2. Check that the heparin syringe is filled.
3. Check for kinks and clamps on the heparin line.
When the pressure in the syringe is within limits again the
heparin pump starts automatically.