HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions B:167
598 UF rate limit is reached
598 UF rate limit is reached
To calculate a new limit adjust time or UF
volume. Check the resulting UF rate.
Appears:
When the actual UF rate differs more than 20 % from the
initially calculated UF rate.
Machine actions:
Alarm sound is delayed for 2 minutes.
Operator action:
Adjust treatment time and/or UF volume to allow the
machine to calculate a new UF rate.
Mute time:
2 minutes.