AK 98 Dialysis Machine - DIASCAN A:139
2) Select the History tab.
3) Press History. The view in Diascan
History will be different if distribution
volume has not been added.
9.6 Set a Kt/V target value
It is possible to set the desired Kt/V target for the patient to have achieved at the end
of the treatment. This value is called Kt/V target.
NOTE!
Before setting a Kt/V target, the dialysis machine needs to know the patient’s
distribution volume. See Section 9.4 ā€œCheck Kt/Vā€ on pageī™›A:137.
Procedure
1) Press the Diascan button.
2) Select the Setup tab.
3) Press Alarm settings.
4) Press Kt/V target.
5) Enter a target value for Kt/V.
6) Press OK.
7) Press Activate and close the menu.