HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:77
CAUTION!
During a functional check the blood line clamps are moving. Be careful and protect
your fingers by not attempting to attach the blood lines until the Blood pump button
is flashing.
Procedure
1) Press the On/Off button.
Check that the alarm indication lights are
activated when the machine is switched
on. See Section 1.1.1 āAlarm indicationā on
pageīB:11 in Alarm handbook.
2) Check that the buttons to the right of the
screen light up when the machine is
switched on. Contact an authorised service
technician if any of the buttons do not light
up. Observe that Blood pump up button
and Blood pump down button do not light
up.
The screen will light up when the machine
is switched on. Contact an authorised
service technician if parts of the screen are
dark and not working.
4.2.3 Set up the dialysis machine
Before you begin
WARNING!
To ensure that the prescription is followed, check that the type of concentrate you
have specified in the dialysis machine is the same as you actually use during
treatment, and that it is the same as is prescribed for the patient.
WARNING!
Check the content of the concentrate container before connecting the pick-up tube
to the concentrate container. If the prescribed concentrate is not used during
treatment this may lead to plasma electrolyte imbalance or acid base imbalance. A
non-intended content may lead to hemolysis.
Procedure
1) Remove the red concentrate connector
from the stand-by port.