B:84 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
509 Blood pump is overloaded
509 Blood pump is overloaded
Check blood lines for obstruction. Start blood
pump.
Appears:
When the blood pump is overloaded and can hardly run in
priming.
Machine actions:
The blood pump stops.
Operator action:
1. Check that the blood pump segment is correctly placed,
see Section 4.2.4 ā€œAttach the arterial blood lineā€ on
pageī™›A:79 in Operators handbook.
2. Press Confirm to continue.
If the attention appears again, contact an authorised
technician.
Mute time:
2 minutes.