A:188 AK 98 Dialysis Machine - Technical data and specifications
HCEN12745 Revision 10.2016
Program version 2.xx
14.6 Environmental data
14.6.1 Operation
If condensation occur when moving the equipment between locations with different
temperatures and high relative humidity (e.g. outdoor and indoor locations), the inside
of the equipment shall be allowed to dry before switching on the equipment.
Ambient Temperature
range
+18 to +35 °C
Relative Humidity
range
15 to 85 % RH
Air Pressure range
(atm. pressure)
700 to 1060 hPa
Table 14-23.
14.6.2 Transportation and storage
It is recommended that the equipment is kept in its original packing during
transportation and storage.
Ambient Temperature
range
-20 to +70 °C
Relative Humidity
range
10 to 93 % RH non-condensing condition.
14.6.3 Electromagnetic environment
The AK 98 dialysis machine is intended for use in the electromagnetic environment
associated with a Professional healthcare facility environment or a Home healthcare
environment as specified below. The customer or the user of the AK 98 dialysis
machine should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The AK 98 dialysis machine uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The AK 98 dialysis machine is suitable for use in
all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A (Not applicable
for 115 V version)
-
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Complies (Not
applicable for 115 V
version)
-

Table of Contents