HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:101
WARNING!
Check that the arterial blood
line is connected to the
rinseback solution to prevent
patient blood loss.
5) Connect the arterial blood line to the rinse
back solution. Open the clamp on the rinse
back line and/or the arterial bloodline.
CAUTION!
Keep an eye on the venous
needle and venous pressure
during the rinse back, since
there is no automatic
supervision at this stage.
6) Press the Blood pump button to start the
blood pump.
7) When the rinse back is finished and the
blood pump stops, disconnect the venous
blood line from the access.
4.3.2 Confirm disconnect patient
WARNING!
To avoid air infusion, ensure that the patient is physically disconnected from the
arterial and venous bloodlines before confirming Disconnect patient. When
Disconnect patient has been confirmed the air detector is deactivated. The
patient must not be reconnected.
Procedure
1) Press Disconnect patient.
2) Confirm that the arterial and venous blood lines are disconnected.
3) Remove the venous blood line from the priming detector.
4.3.3 Machine aftercare
Before you begin
CAUTION!
To ensure that the patient is not connected when doing machine aftercare, check
that the arterial and venous blood lines have been completely disconnected from the
patient.
Procedure
1) Turn the dialyzer so that the blue
connectors are up.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals