AK 98 Dialysis Machine - Attentions B:107
533 Disinfection required
533 Disinfection requiredī¼
{0} days since last disinfection. To clear
attention press Confirm.ī¼
Appears:
When a disinfection program is needed.
The period of time between the last performed disinfection
and the appearance of the attention can be preset.
Machine actions:
None.
Operator action:
If possible, run a heat disinfection program.
If the time or situation is not suitable for a heat disinfection
program, press Confirm to clear the attention. Disinfect the
dialysis machine after the treatment, or the attention will
appear again during the next treatment.
Mute time:
Permanent.