AK 98 Dialysis Machine - Attentions B:73
2.1 Attention indication
When an attention is generated, the machine indicates an
attention as follows:
The Information tab is flashing.
The Information tab turns blue
and the attention message is
displayed.
The indication light above the
screen shows a flashing blue
light (frequency 1 Hz).
-
-
-
(5 sec)
-
-
-
(5 sec)... For most attentions the buzzer
sounds. The buzzer sound for
attentions is called notification
sound. The notification sound is
a recurring sound of three tones
that are repeated with a 5
second intermission.
The Mute button is flashing.
The light stops flashing when the attention is confirmed
and/or the cause of the attention is corrected.
2.2 General attention handling
Press the Information tab.
The attention message is displayed. The Mute button lights
up and the buzzer is muted (the mute time varies
depending on the attention).