AK 98 Dialysis Machine - Alarms B:21
104 Blood is not detected
104 Blood is not detectedī¼
Check cause and press Confirm.ī¼
Appears:
When the connect patient procedure has been started, the
connect patient volume is achieved and no blood has been
detected in the venous blood line.
Machine actions:
ā The blood pump stops.
ā The venous blood line clamp closes.
Operator action:
1. Check that the venous blood line is properly inserted
into the priming detector.
2. When the priming detector detects blood, the blood
path on the screen lights up red and treatment time
starts to count down.
3. Press Confirm.
4. Select Continue connect patient and press
Confirm.