B:110 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
536 Fluid path obstruction
536 Fluid path obstruction
Too high blood circuit pressure. Check
circuit, start blood pump.
Appears:
When the machine detects an obstruction in the
extracorporeal circuit (blood lines and dialyzer) or in the
fluid path.
Machine actions:
The blood pump stops.
Operator action:
Check for obstructions in the extracorporeal circuit, for
example kinked blood lines or clotted dialyzer and/or blood
lines. Restart the blood pump when the obstruction is
corrected.
Mute time:
2 minutes.