B:158 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
588 Set fluid flow is not reached
588 Set fluid flow is not reached
Adjust dialysis fluid flow or call technical
service.
Appears:
The actual dialysis fluid flow rate is not the same as the set
value. The dialysis machine cannot set the required dialysis
fluid flow rate.
Machine actions:
None.
Operator action:
If the treatment can continue with a lower flow, adjust the
set value to the actual value.
If this is not possible, contact an authorised technician.
Mute time:
Permanent.