A:100 AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
2) Press the flashing Blood pump button to
restart the blood pump.
3) Adjust the blood flow rate using the Blood
pump up and Blood pump down buttons.
4) Press the flashing Ultrafiltration
button to start ultrafiltration.
5) Press the flashing Arterial pressure
control to centralise the alarm limits around
the actual value.
6) Press the flashing Venous pressure control
to centralise the alarm limits around the
actual value.
7) Check that the venous drip chamber is
filled to the correct level, and that the short
lines on the top of the chamber are
clamped.
4.3 End a double needle treatment
4.3.1 End the treatment
When the treatment is finished and treatment time on the screen has counted down to
0:00, an attention will appear.
596 Treatment time expiredī¼
To discontinue treatment press Confirm.ī¼
NOTE!
If the remaining treatment time is not already 0:00, press the Time indicator and set
Treatment time to 0:00, to access the normal end the treatment procedure.
NOTE!
To interrupt the treatment, press the Time indicator and decrease the treatment time
to 0:0.
Procedure
1) Press the attention to confirm end of
treatment.
2) Press the Rinse back button that appears
in the Information field.
3) Press Confirm. The blood pump stops.
4) Disconnect the arterial blood line from the
access.