A:94 AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
10) Only for single needle treatment: The
expansion chamber is filled to a suitable
level during priming. Close the clamp
above the expansion chamber. See
Section 5 āHaemodialysis - Single needle
treatmentā on pageīA:105 for more
information.
4.2.8 Priming options
4.2.8.1 Extra priming
When the priming is complete an information message will be displayed, containing a
selection of possible actions.
Procedure
1) Check that you have enough priming fluid.
2) Press Extra priming.
3) Press the flashing Blood pump button.
4)
CAUTION!
Check carefully that the
venous blood line is correctly
placed in the priming
detector to make sure that
the supervision of alarms is
activated.
When the extra priming is completed:
Press Connect patient.
5) Continue starting the treatment. See
Section 4.2.9 āSet treatment timeā on
4.2.8.2 Recirculation
When the priming is complete an information message will be displayed, containing
the possible actions to recirculate.
You can set the priming fluid to circulate in the blood lines.
Procedure
1) Press Recirculation.
2) Connect the two patient ends of the blood lines to make a closed circuit. Use a
sterile disposable blood line connector.