EasyManuals Logo

Gambro AK 98 Operator's Manual

Gambro AK 98
386 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #261 background imageLoading...
Page #261 background image
HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Alarms B:61
218 TMP is too low
218 TMP is too low
Appears:
When the transmembrane pressure (TMP) has reached the
set low alarm limit.
Machine actions:
None.
Operator action:
1. Press the Fluid button.
2. Select the TMP tab.
3. Check for a cause: unexpected patient weight loss or
issues with the needle, extracorporeal circuit or dialyzer
(clogged dialyzer), extracorporeal circuit kink or clamp
on the venous blood line, and take appropriate actions.
4. Do standard patient checks, for example measure
blood pressure.
5. If possible, decrease the set treatment time, increase
the set UF volume or decrease the blood flow rate.
6. If possible, adjust the low TMP alarm limit.
7. Check that the blood line between the dialyzer and the
venous drip chamber is not clamped.
8. Confirm that the actual TMP value and alarm limits are
in accordance with the actual UF rate and UF
coefficient.
9. If necessary, end the treatment and call an authorised
technician.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals