HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions B:97
522 Decalcification required
522 Decalcification requiredī¼
{0} treatments since last decalcification. To
clear attention press Confirm.ī¼
Appears:
When a decalcification program is needed. Too many
treatments performed since last decalcification.
The number of treatments between the last performed
decalcification and the appearance of the attention can be
preset.
Machine actions:
None.
Operator action:
If possible, run a decalcification program, for example a
heat disinfection program with CLEANCART C cartridge or
liquid citric acid. (A Short heat Citric program cannot be
used).
If the time or situation is not suitable for a decalcification
program, press Confirm to clear the alarm. Disinfect the
dialysis machine after the treatment, or the attention will
appear again during the next treatment.
Mute time:
Permanent.