B:76 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
501 Air leakage in fluid path
501 Air leakage in fluid pathī¼
Check fluid tube connectors to dialyzer.ī¼
Appears:
When the dialysis fluid tubes are not properly connected to
the dialyzer.
Machine actions:
None.
Operator action:
Check the dialysis fluid tubes connections. See Section
4.2.7 āPriming the dialysis circuitā on pageīA:91 in
Operators handbook.
Mute time:
2 minutes.
WARNING!
If no action is taken for this attention, the patient may get
incorrect weight loss during the treatment.