AK 98 Dialysis Machine - Alarms B:45
202 Fluid tubes in safety couplings
202 Fluid tubes in safety couplingsī¼
Connect the fluid tubes to the dialyzer.ī¼
Appears:
ā When blood is detected by the priming detector when
connecting the patient.
ā When the dialysis fluid tubes are connected to the
safety couplings during treatment.
Machine actions:
None.
Operator action:
Move the dialysis fluid tubes from the safety couplings to
the dialyzer.