AK 98 Dialysis Machine - Alarms B:25
108 Fluid path obstruction
108 Fluid path obstructionī¼
Too high blood circuit pressure. Check
circuit, start blood pump.ī¼
Appears:
When the fluid path is blocked.
Machine actions:
The blood pump stops.
Operator action:
1. Press the Fluid bypass button.
If the alarm disappears, the obstruction is external, for
example located in the blood lines, the dialysis fluid
tubes or the dialyzer.
2. Check for obstructions in the blood lines, the dialysis
fluid tubes or the dialyzer and remove them if you find
any.
3. Press the Fluid bypass button.
If the alarm does not disappear, the obstruction is internal.
Call an authorised technician.