HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions B:101
526 Diascan clearance is too low
526 Diascan clearance is too low
Check treatment settings and check/adjust
clearance limit. To continue press Confirm.
Appears:
When the measured clearance value is below the set alarm
limit.
Machine actions:
None.
Operator action:
1. Check for a cause.
2. Adjust treatment parameters to increase clearance or
adjust the alarm limit for clearance.
Mute time:
Permanent.