AK 98 Dialysis Machine - Attentions B:123
549 Functional check is stopped
549 Functional check is stopped
To continue, disconnect arterial pressure
transducer connector.
Appears:
When the arterial pressure transducer of the arterial blood
line has been attached too early in functional check. See
Section 4.2.4 ā€œAttach the arterial blood lineā€ on pageī™›A:79
in Operators handbook.
Machine actions:
None.
Operator action:
Disconnect the arterial pressure transducer. If the attention
appears again, contact an authorised technician.
Mute time:
5 minutes.