B:94 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
519 Cleaning is required
519 Cleaning is required
{0} hours since last cleaning. To clear
attention press Confirm.
Appears:
When a cleaning program is needed.
The period of time between the last performed cleaning and
the appearance of the attention can be preset.
Machine actions:
None.
Operator action:
If possible, run a cleaning program, for example a heat
disinfection program with CLEANCART A cartridge or a
chemical disinfection program with sodium hypochlorite.
If the time or situation is not suitable for a cleaning program,
press Confirm to clear the attention. Clean the dialysis
machine after the treatment, or the attention will appear
again during the next treatment.
Mute time:
Permanent.