AK 98 Dialysis Machine - Attentions B:137
563 Improper time and/or UF volume settings
563 Improper time and/or UF volume settings
Adjust time and/or UF volume.
Appears:
When a too high UF volume has been set in combination
with the set treatment time.
Machine actions:
After 2 minutes there is a buzzer sound.
Operator action:
Adjust Time (Isolated UF time), or UF volume (UF
volume), until the calculated UF rate is lower than the high
limit for UF rate.
Mute time:
2 minutes.