B:140 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
566 Incorrect conductivity
566 Incorrect conductivity
Check set values and concentrate/connectors
and press Confirm.
Appears:
When the conductivity is not correct during a functional
check.
Machine actions:
None.
Operator action:
1. Check that the red concentrate connector is properly
connected to the pick-up tube and the concentrate
container, see Section 4.2.3 ā€œSet up the dialysis
machineā€ on pageī™›A:77 in Operators handbook.
2. Check that the correct concentrate is connected.
3. Check that the SoftPac container pin is broken.
4. Press Confirm.
Mute time:
5 minutes.