A:84 AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
10) Connect the arterial blood line with the
saline or priming solution bag. Check that
the connections are tight.
11) Attach the arterial pressure transducer
protector of the arterial blood line to the red
arterial pressure transducer connector.
Press to pierce the membrane and then
turn to secure the connection.
12) Close all relevant clamps.
CAUTION!
Check that all relevant
clamps are closed to prevent
undetected blood loss.
4.2.5 Attach the venous blood line
The venous blood line has blue connectors.

Table of Contents