A:174 AK 98 Dialysis Machine - Maintenance handling
HCEN12745 Revision 10.2016
Program version 2.xx
Every week, disinfect the pick-up tubes by wiping the outside with 70% ethanol and
flushing the inside with 70% ethanol. Let them dry naturally ā do not wipe them.
13.6 Surface
After each treatment, you need to clean and disinfect the exterior surface and the top
tray of the dialysis machine. Wipe them with a cloth moistened with ethanol (70 %) or
isopropanol (60 %).
CAUTION!
In order to protect the internal parts of the AK 98 dialysis machine against spillage,
the top tray shall always be placed on top of the machine, except during technical
service.
CAUTION!
The leakage detector sensor and leakage detector tray shall on a regular basis be
cleaned and inspected for cracks.
NOTE!
Do not use disinfectants containing tensides or iodine-based disinfectants. Such
disinfectants may discolour and cause cracks on the surface of the dialysis machine
cabinet.
13.7 Change ultrafilter
WARNING!
Check carefully that there is no leakage from the ultrafilter after changing it.
WARNING!
Before a new treatment can be performed after changing the ultrafilter, the machine
must go through a disinfection to ensure the quality of the dialysis fluid.
NOTE!
The ultrafilter is not available on all product configurations of the AK 98 dialysis
machine.
Replace the ultrafilter regularly, to maintain
high hygienic quality.
NOTE!
Make sure that the ultrafilter is
handled in an aseptic way
according to the corresponding
package insert.
Procedure
1) Pull the handle of the lower latch to open
the locking mechanism and press the
lower latch downwards.