B:184 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
617 Leakage detected during disinfection
617 Leakage detected during disinfection
Internal fluid leakage detected. Disinfection
is aborted. Not possible to continue
disinfection.
Appears:
When the machine has detected a fluid leakage during
disinfection.
Machine actions:
ā— Close valves to stop leakage to the environment.
ā— Disinfection is stopped.
Operator action:
1. Empty the fluid leakage detector.
2. Contact an authorised technician.
It is not possible to continue disinfection.
Mute time:
2 min