A:146 AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
NOTE!
For the best cleaning results, start with a decalcification.
10.3.3 Start a heat disinfection
Procedure
1) Press the Disinfection button.
2) Select the Heat tab.
3) Press Heat.
4) Press Confirm.
10.3.4 Start a heat disinfection with a CLEANCART cartridge
Procedure
1) Press the Disinfection button.
2) Select the Heat tab.
3) Press Heat CleanCart.
4) Press Confirm.
5) When you are prompted by an attention:
Attach the CleanCart cartridge to the
holder.
10.3.5 Start a heat disinfection with liquid citric acid
Procedure
1) Press the Disinfection button.
2) Select the Heat tab.
3) Press Citric 20%.
4) Make sure that the disinfectant inlet tube is
placed in the disinfectant solution.