AK 98 Dialysis Machine - Alarms B:59
216 Heparin pump technical error
216 Heparin pump technical error
Heparin pump failure. To confirm not using
heparin pump, set heparin values to zero.
Appears:
When there is a technical fault with the heparin pump.
Machine actions:
The machine actions differ depending on the kind of
technical error.
Operator action:
1. Set the heparin values to zero to confirm that the
heparin pump is not going to be used:
- Press the Blood button.
- Select the Heparin tab.
- Set the heparin values to zero.