AK 98 Dialysis Machine - Attentions B:141
567 Incorrect dialysis fluid composition
567 Incorrect dialysis fluid compositionī¼
Check selection of concentrate.ī¼
Appears:
When the dialysis fluid composition deviation (the relation
between the acidic and the bicarbonate concentrate) is too
high.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
Check what concentrate is set and what is connected.
Make sure they are the same. See Section 4.2.3 āSet up
the dialysis machineā on pageīA:77 in Operators
handbook.
Mute time:
2 minutes.