B:74 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
2.3 How to find the attention
The attention descriptions in this chapter are arranged in
numerical order. Each attention description has the
following sections:
Appears Describes what conditions must
be present for an attention to
appear; what triggers/generates
the attention.
Machine actions Describes which actions that
will automatically be taken by
the machine due to the
attention.
Operator action Describes the actions that are
expected from the operator
when an attention appears.
Mute time When the mute button is
pressed, the buzzer sound will
be muted during the specified
period of time
When {0} is displayed in the descriptive text in an attention,
it is replaced with a live integer value.
2.4 List of attentions
500 Air detector not activated
500 Air detector not activatedī¼
Activate air detector.ī¼
Appears:
During priming, 15 seconds after the venous drip chamber
has been filled upon another attention request.
Machine actions:
None.
Operator action:
Press the Air detector button to activate the air detector
alarm function.