B:156 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
585 Profiling is not restarted
585 Profiling is not restartedī¼
Check profiling and UF, Naāŗ and HCO
3
+
set
values and reactivate profiling.ī¼
Appears:
When treatment is resumed after the dialyzer and the blood
lines have been changed during treatment.
Machine actions:
The machine runs with the conductivity and/or the UF rate
settings present when New blood circuit was selected
and discontinuing mode was activated.
Operator action:
Check actual set profiling values and reactivate profiles.
1. Press the Fluid button.
2. Check that the start and stop values for one or both of
the following profiling type are correct.
- Select the UF rate tab and then press Profiling.
- Select the Cond tab and then press Profiling.
3. Reactivate the profiling function, see Section 7
āProfilingā on pageīA:117 in Operators handbook.
Mute time:
2 minutes.