B:176 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
607 Wrong disinfectant
607 Wrong disinfectant
Check disinfectant. To continue press
Confirm.
Appears:
When the dialysis machine detects a too high or too low
disinfectant concentration.
The appearance of the attention can be preset.
Machine actions:
None.
Operator action:
1. Check that the disinfectant set in the dialysis machine is
the same as the actual disinfectant connected to the
machine.
2. Press Confirm to repeat the fill-up phase.
Mute time:
Permanent.