B:86 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
511 Blood pump test failed
511 Blood pump test failedī¼
Check pump segment and arterial bloodline for
obstruction. To continue press Confirm.ī¼
Appears:
When the blood pump rotation test fails.
Machine actions:
The machine stops the blood pump test until the attention is
confirmed.
Operator action:
Check the blood line and confirm the attention.
Mute time:
2 minutes.