B:180 AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
613 Treatment history was not transferred to the server
613 Treatment history was not transferred to
the server
Document treatment history manually. To close
this message, press Confirm.
Appears:
If the machine has a permanent failures sending treatment
data to the Clinical Information System.
Appears when the operator presses the Disinfection
button and unsent messages remain or permanent failures
have occurred.
Machine actions:
None
Operator action:
Treatment data in the Clinical Information System is not
complete. Treatment information is available in the History
tab.
Manual documentation on paper, etc. is needed.
1. Press Confirm to continue.
Mute time:
Permanent