AK 98 Dialysis Machine - Attentions B:181
614 Leakage detected
614 Leakage detected
Internal fluid leakage detected. To continue,
wipe fluid detector container dry.
Appears:
When the machine has detected a fluid leakage.
Appears in:
ā— Function check
ā— Blood line preparation
ā— Pre treatment
ā— Post treatment
Machine actions:
ā— Close valves to stop leakage to the environment.
Operator action:
1. Wipe the fluid leakage detector tray dry.
2. Continue treatment.
Mute time:
2 min